Abstract
Oversulfated chrondroitin sulfate (OSCS) was recently identified as a contaminant of heparin and was associ- ated with serious adverse events in patients treated with heparin. Because heparin is a common component of blood col- lection tubes, we tested the effect of OSCS on the laboratory analysis of plasma. Blood from healthy volunteers (N=50) was collected into tubes containing various mixtures of heparin and OSCS. Samples were inspected for microclots and were analyzed for a panel of 28 routine laboratory tests. No microclots were observed in tubes that contained only heparin but were detected in 18%, 88% and 76% of plasma samples containing 5%, 15%, 20% OSCS (%weight relative to hepa- rin), respectively. OSCS at the highest dose (20%) caused a systematic bias for the following 6 tests: Lactate Dehydro- genase: 18% (12% to 24%); Triiodothyronine: -5.7% (-8.1% to -3.3%); Potassium: -2.8% (-4.2% to -1.4%); Total Protein: 2.5% (1.4% to 3.6%); Chloride: -1.4% (-1.8% to -1.0%) and Uric Acid: 1% (0.5% to 1.4%). In summary, OSCS contami- nation of heparin was found to potentially affect the anticoagulation of plasma and the analytical performance of several routine clinical laboratory tests.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.