Abstract

Supercritical fluid chromatography of PAHs was performed with pure carbon dioxide and helium headspace carbon dioxide at various cylinder fill levels. The retention times of the PAHs increased when helium headspace carbon dioxide was used as a carrier fluid relative to pure carbon dioxide. The increased retention times were affected by the level of the liquid phase present in the helium headspace carbon dioxide cylinder. As more liquid phase was removed from the cylinder, the effect of helium on the solvating power of CO(2) was reduced because the relative amount of helium dissolved in the liquid phase decreased. Furthermore, the effect of helium headspace carbon dioxide was investigated with methanol-modified carbon dioxide mobile phases for the analysis of steroids. We observed that the relative solubility of helium in carbon dioxide resulted in longer retention times when compared to pure carbon dioxide as the liquid level of carbon dioxide decreased.