Abstract

Objectives: The analytical performance of the TOSOH HLC-723G7 hemoglobin HPLC analyzer and the effect of the presence of HbS in the determination of HbA 2 using HPLC and manual column methods. Design and methods: The performance characteristics of the TOSOH HLC-723G7 analyzer in the determination of HbA 2 were compared to those of the HELENA Beta-Thal Quik column. The effect of HbS presence in the samples was quantified using the HELENA SAS-MX alkaline gel electrophoresis kit as the reference method. Results: Within-run and between-run CVs for HbA 2 were better for the TOSOH HPLC analyzer than for the HELENA manual column method. The presence of HbS in the samples produces a strong positive bias in the %HbA 2 values when using both the HPLC and manual column methods, compared to the alkaline electrophoresis gel. Conclusion: Both the TOSOH HPLC and the manual column are reliable methods for %HbA 2 determination when no HbS is detectable in the samples. When HbS is present, the gel electrophoresis method gives more accurate results.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.