Abstract
The effect of interposing a half-wavelength membrane between the transducer and the load in mechanically sectored ultrasonic scanners was investigated. In particular, the effects of impedance mismatch and detuning on broadband spectra were theoretically and experimentally studied. These studies indicate that for certain common cases, a bandwidth reduction of 20%-30% can be anticipated, resulting in a proportional decrease in the axial resolution.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.