Abstract

The effect of glucose content on the electrochemical corrosion behavior of the Ti/ZrO2 brazing joint in simulated body fluid (SBF) was researched by the means of SEM morphologies, electrochemical and XPS analyses. Herein, pitting is observed to be a dominating corrosion model under the investigated glucose content. The pitting corrosion of the joint in 200 mg/dL SBF is minimal. In addition, the joint in 200 mg/dL SBF manifests the best corrosion resistance by electrochemical analyses, which indicates that glucose content has a bidirectional effect on corrosion of the Ti/ZrO2 brazing joint. Additionally, the corrosion current value and impedance of titanium and brazing joint are close, which indicates that their corrosion resistance is similar. Finally, the OH-, Cl-, Sn2+/Sn4+ and -COOH on the joint surface are found by XPS analysis, and the mechanism of Ti/ZrO2 brazing joint corrosion is elucidated. The study provides a novel understanding of the corrosion behavior and relevant corrosion mechanism of the Ti/ZrO2 brazing joint in body fluids with different glucose content.