Abstract

Backgound: The effect of anesthetic techniques, such as closed circuit anesthesia (CCA) using semiclosed circuit system and semiclosed circuit anesthesia (SCCA), on the work of breathing has not been studied yet in detail. This study was purposed to compare the work of breathing according to anesthetic technique (CCA, SCCA). Methods: Thirty patients were assigned to receive either SCCA group or CCA group (n = 15). Anesthesia was induced with propofol 2 mg/kg with 2% lidocaine 1 ml. Two percents isoflurane with and 2 L/min were given for 10 min to patients initially to wash in functional residual capacity and the breathing circuits. In SCCA group, anesthesia was maintained with 2% isoflurane in 2 L/min and 2 L/min throughout the surgery. In CCA group, was reduced to 200 ml/min and to 100 ml/min with isoflurane vaporizer setting adjusted to 4% for anesthesia maintenance. When the operation was ended, the vaporizer setting of isoflurane deceased to zero and then was increased to 4 L/min for the arousal of the patient. We measured the inspiratory/expiratory concentration of isoflurane, end-tidal C, the hemodynamic parameters, the change of airway pressure, the work of breathing, and compliance at anesthetic induction and emergence in both groups. Results: There were no significant differences in the inspiratory/expiratory concentrations of isoflurane, the hemodynamic parameters, end-tidal C, airway pressure, the work of breathing and compliance between the groups. Conclusions: CCA using semiclosed circuit system does not increase the work of breathing compared to SCCA.