Abstract
We investigated the behavior of speckle contrast and size under various experimental conditions using 1.82 keV x-rays. In this paper, we report the comparison of two different setups for x-ray speckle experiments: one employing a focusing zone plate and one in which a pinhole selects the size of the coherent x-ray beam. We found a strong dependence of the speckle contrast and size on the type of setup. In general, the pinhole setup results in higher contrast but smaller speckle size. On the other hand, the zone plate setup allows one to target much smaller areas of interest in the sample, down to submicron dimensions, and also to adjust the speckle size. We anticipate that these results will be useful in future time-correlation spectroscopy experiments.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.