Abstract

The effect of defocusing on the quality of signals from live cells detected by an automated image cytometry device, the Cell Analyzer, was examined. The influence of these effects on the ability of this device to automatically locate cells plated into a tissue culture flask was then determined by measuring the performance of cell detection and recognition procedures as a function of focus setting. Acceptable limits for deviation from the optimal focus setting (as determined by microscope objective position) were found to be similar for both these procedures, ranging from 40 microns below to 25 microns above the optimal focus position. These limits were asymmetrical about ideal focus due to a pronounced asymmetry in the effects of positive and negative defocusing on the cell signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.