Abstract
Novel methods are presented which allow the estimation of the effect of stereotactic fiducial marker localization on target localization in a stereotactic surgery planning program based on the integration of image data from various modalities. The results of simulations are presented. A method for the evaluation of the effect of fiducial marker localization error on stereotactic localization, both for computed tomography (CT) and digitized radiography. Is proposed based on perturbation studies. For tomographic data, the comparison of the stability of two types of fiducial localizers: a standard commercial device based on nine (six vertical and three oblique) rods and an imaginary system with an extra set of rods, called '4-N', is discussed. Results of the perturbation studies clearly indicate the necessity of sub-pixel registration of the fiducial markers in CT. On the other hand, results demonstrate that stereotactic localization based on projection images is less sensitive to fiducial localization error than CT. Finally, the authors present an algorithm for the localization of fiducial markers to sub-pixel accuracy in radiographs acquired with a commercial angio localizer.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.