Abstract

To ascertain the effects of exposure parameters (tube current and tube voltage) and the gutta-percha cone (GPC) size on root fracture-like artifacts obtained with cone-beam computed tomography (CBCT). Fracture-like artifacts appearing on CBCT images of nine extracted human mandibular premolars filled with GPCs of size #50 or #80 were analyzed using six exposure factors: two tube voltages (80kV and 110kV); and three tube currents (4mA, 7mA, and 10mA). On axial images, the gray value (GV) was recorded at three points: the mesiobuccal portion (MBP) as the sound dentin, the mesial portion (MP) as the artifact line, and the water area (WA). The rate of decrease in the GV (RDGV) of the artifact line was calculated using the formula: RDGV (%)=(GV of MBP-GV of MP)×100/(GV of MBP-GV of WA). Comparison of the #80 group and the #50 group with equal tube voltages and tube currents shows that artifact lines in the #80 group were more obvious than those in the #50 group. The artifact lines with 80 kV were markedly more visible than those with 110 kV for each tube current and GPC size. Tube current changes did not affect the artifact line for any tube voltage or GPC size. For the reduction of artifacts, we recommend selection of higher tube voltages and lower tube currents when taking CBCT images of teeth with each GPC size.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call