Abstract
C80 micro calorimeter was applied to investigate the initial heat release temperature and heat output at low-temperature oxidation process of Kabuliang anthracite coal under the experimental conditions of different coal sample mass, different temperature rising rates and different gas flow, and analyzed the effect of experimental conditions on test results. The results indicate that the coal sample mass and temperature rising rate affect measurement results, while there is no effect of gas flow. Under the same experimental condition, a higher temperature rising rate leads to a higher initial heat release temperature and less output of heat; The initial heat release temperature shows the trend of increase after decrease with the increase of coal sample mass. So, when micro calorimeter is used to investigate the coal low-temperature oxidation process, the experiments under the conditions of appropriate coal sample mass and the lower temperature rising rate can obtain more accurate test parameters.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.