Abstract

Glucose oxidase (β- d-glucose:oxygen 1-oxidoreductase, EC 1.1.3.4) was immobilized in a crosslinked matrix of bovine serum albumin, catalase, glucose oxidase and glutaraldehyde on platinum foil. When placed in glucose soltuion, this enzyme-electrode elicited a potentiometric response that varied with the changes in glucose concentration. The immobilized glucose oxidase was present at 7.4–10.1 μg enzyme proten/ml of matrix, as determined with 125I-labelled enzyme. The coupled enzyme activity was stable over 120 h; however, the apparent activity of the immobilized glucose oxidase was markedly less than that for the same amount of enzyme free in solution. This indicated a significant level of diffusional resistance within the enzyme-matrix. The potentiometric response to glucose increased significantly as either the thickness of the enzyme-matrix or the glutaraldehyde content was reduced; this also was attributed to diffusional effects. Several enzyme-electrodes, constructed without exogenous catalase and with different amounts of glucose oxidase, showed greater sensitivity in potentiometric response at low glucose oxidase loadings. These results are consistent with the hypothesis that the potentiometric response arises from an interfacial reaction involving a hydrogen peroxide redox couple at a platinum surface. The data also suggest that an optimum range of hydrogen proxide concentration exists for maximum electrode sensitivity.