Abstract
A system is described which allows operation of a range of capillary based liquid phase separations including capillary electrophoresis, isocratic and gradient capillary electrochromatography, isocratic and gradient capillary liquid chromatography and electrically assisted gradient capillary liquid chromatography. The system was coupled to electrospray ionization mass spectrometry in the electrically assisted capillary liquid chromatography mode to investigate the effect of applied voltage on the selectivity in peptide mapping separations. Analyses were performed on tryptic digests of recombinant human growth hormone and tissue plasminogen activator. The results show a small but useful effect on selectivity that can be used to fine tune specific separations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.