Abstract

Although several clinical studies on retrieved tibial bearings have been published so far, monitoring and comparing the in vivo performance of different types of UHMWPE is still a much needed task. The present study reports for the first time results on the effect of sterilization by electron beam on the mid-term in vivo performance of conventional UHMWPE tibial plates. In the present investigation, visual inspection of wear damage based on the Hood's scoring method, Raman micro-spectroscopy and Fourier-transformed infrared spectroscopy were utilized to unveil the damage, the microstructural modifications and the oxidation occurred during implantation. The findings of this investigation have been discussed and compared to previous clinical studies on γ-air sterilized, γ-inert sterilized tibial bearings.