Abstract

OBJECTIVES To investigate the effect of dose reduction in digital panoramic radiography on subjective image quality and diagnostic performance. METHODS Two digital panoramic radiographs were obtained with the Orthophos DS(Sirona, Bensheim, Germany) of patients (n=100) receiving dental care. The first image was taken at the standard exposure setting. For the second image the tube current was reduced by between 48 and 53%, 63 and 69%, 75 and 80% and 80 and 81% respectively. Ten observers rated both images for 21 specific anatomical details and 30 pathological findings. RESULTS All radiographs taken at reduced mA levels had a significantly inferior score (P<0.01) for anatomical details. There was no difference in the scores for pathological findings. CONCLUSION Radiographs obtained at reduced mA had inferior subjective image quality, but there was no difference in diagnostic performance. Thus, a reduction in tube current of approximately 50% is recommended. In certain circumstances such as follow-up radiographic examinations, a reduction of up to 65% should be considered.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.