Abstract

The new TL LiF:Mg,Cu,Ag material was prepared and investigated in this study. The TL intensity of LiF:Mg,Cu,Ag is strongly dependent on the concentration of dopants and the preparation procedure. Any small change in these factors can cause alterations in TL response. In this study, the influence of Cu and Ag concentrations on the response of the LiF:Mg,Cu,Ag sample was investigated and showed that the height of the low, main and high temperature peaks changes with Ag concentration. Their intensities increased with increasing Ag concentration to a maximum value and decreased with higher Ag concentration. It was also found that Cu concentration less than 0.05 mol% influences the maximum peak height and TL intensity. The optimum Cu and Ag concentrations were found to be 0.05 and 0.1 mol% at 1005 °C QT, respectively. The role of dopants in LiF:Mg,Cu,Ag material was also investigated. The results showed that presence of three dopants is important for having material with sensitivity higher than LiF:Mg,Ti. The Mg dopant plays a crucial role in the formation of the trapping center and the position of the main dosimetric peak.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.