Abstract

The normal functioning of dual chamber pacemaker-cardioverter defibrillator (AV pacer/ICD) may be affected by oversensing of the farfield R wave (FFRW) by the atrial channel. This study aimed to investigate whether placement of the AV pacer/ICD's atrial lead at a lateral (LAT) wall location compared to a medial (MED) location i.e. the appendage of the right atrium, would reduce the amplitude of FFRWs but not the nearfield atrial electrograms (AEGMs) during sinus rhythm (SR) and ventricular fibrillation (VF). In 17 patients, real time electrograms were recorded during SR and induced VF through the atrial lead initially at the MED and subsequently at the LAT location. In 10 patients the electrograms in SR were also recorded on a computerized data acquisition and recording system at different band-pass filter settings. Although FFRWs were recorded both at MED and LAT locations, they were much smaller, 3.5+/-4.1mm during SR and 1.7+/-2.2mm during VF at the LAT location. At 30-500Hz band-pass filter, lower amplitudes of FFRWs 0.14+/-0.09 mV were recorded at the LAT location. The V/A ratios of the amplitudes of FFRWs and AEGMs were smaller at the LAT location during SR and VF. The nearfield AEGMs were of similar amplitudes at the MED and LAT locations. These data indicate that lower amplitudes of FFRWs are recorded by placement of the atrial lead at the lateral wall of the right atrium. Oversensing of FFRWs may be prevented to improve functioning of the AV pacer-ICD.