We examine the effects of detector size on the imaging properties of a scanning phase contrast microscope. The lateral size of the point-spread function is found to be wider when a point detector is used than when a much larger detector is used. Combining this finding with the well-known difficulties involved in using a confocal microscope in transmission mode suggests that a phase contrast microscope employing a point detector will in general have inferior imaging properties compared with one employing a large detector.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.