Abstract

Background: Phantom studies are widely used to assess variability in measurements. This study aimed to assess the reliability and accuracy of point Shear Wave elastography (pSWE) measurements of an elasticity phantom. Methods: Measurements were obtained by an experienced certified clinical sonographer at three different depth levels in kPa, using a curvilinear 5-1MHz transducer of the EPIQ7 ultrasound imaging system. Results: A total of 180 pSWE measurements were obtained at three different depth levels (three cm, five cm, and seven cm) of the phantom background. The mean CV of pSWE was low at all depths (3 cm: 8.8%; 5 cm: 7%; 7 cm: 7.2%). There was a significant difference between measurements at depths of 3 cm vs. 7 cm (MD: −0.85, 95% CI −1.5, −0.11, p = 0.024) and measurements at depths 5 cm vs. 7 cm (MD: −1.1, 95% CI −1.7, −0.47, p = 0.001). An overestimation of mean pSWE measurements at a depth of 7 cm was noted compared to the manufacturer’s value (2.7%, p = 0.006). Conclusions: Superficial phantom SWE measurements in this study had low variability compared to deep measurement. pSWE measurements at deep levels can be overestimated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.