To test longevity of image quality in storage phosphor plates (SPPs) at various exposure settings, storage conditions, and delays in scanning. Fifteen Digora plates were exposed from 0.08 to 0.20 seconds and scanned immediately, 10, 30, and 60 minutes, and 24 hours after exposure. Plates were stored both in daylight and in a light-tight box. Mean gray values (MGVs) were compared using 2 x 5 x 5 factorial ANOVA. Interaction between variables was tested using Bonferroni/Dunn multiple comparisons test. MGVs decreased with increase in exposure but increased with the scan delay. Only MGVs of plates scanned within 10 minutes after exposure were not significantly different from the ones scanned immediately ( P > .05). MGVs increased with scan delay for all exposure times no matter how the plates were stored ( P < .05). Based on the time delays examined, it is recommended to scan the Digora SPP no later than 10 minutes after exposure. Longer periods may cause loss of quality.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.