Abstract

We aimed to evaluate the effect of tube voltage, tube current-time product, and iterative reconstruction on iodine quantification using a dual-layer spectral CT scanner. Two mediastinal iodine phantoms, each containing six tubes of different iodine concentrations (0, 1, 2.5, 5, 10, and 20 mg I/mL; the two phantoms had tubes with contrast media diluted in water and in 10% amino acid solution, respectively), were inserted into an anthropomorphic chest phantom and scanned with varying acquisition parameters (120 and 140 kVp; 20, 40, 60, 80, 100, 150, and 200 mAs; and spectral reconstruction levels 0 and 6). Thereafter, iodine density was measured (in milligrams of iodine per milliliter) using a dedicated software program, and the effect of acquisition parameters on iodine density and on its relative measurement error (RME) was analyzed using a linear mixed-effects model. Tube voltages (all, p < 0.001) and tube current-time products (p < 0.05, depending on the interaction terms for iodine density; p = 0.023 for RME) had statistically significant effects on iodine density and RME. However, the magnitude of their effects was minimal. That is, estimated differences between tube voltage settings ranged from 0 to 0.8 mg I/mL for iodine density and from 1.0% to 4.2% for RME. For tube current-time product, alteration of 100 mAs caused changes in iodine density and RME of approximately 0.1 mg I/mL and 0.6%, respectively. Spectral level was not an affecting factor for iodine quantification (p = 0.647 for iodine density and 0.813 for RME). Iodine quantification using dual-layer spectral CT was feasible irrespective of CT acquisition parameters because their effects on iodine density and RME were minimal.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call