Abstract

Purpose: To investigate the effects of central corneal thickness (CCT) on intraocular pressure (IOP) measurements of the Pascal dynamic contour tonometry (DCT), Canon TX‐10 non‐contact tonometry (NCT) and Goldmann applanation tonometry measurements (GAT) in healthy subjects.Methods: IOP values of 135 eyes with normal corneas of 135 healthy volunteers were determined by DCT, NCT and by GAT. The CCT was measured using an ultrasonic pachymeter after all IOP determinations had been made.Results: When DCT measurements were compared (IOP = 17.52 ± 2.0-mmHg) with NCT measurements (IOP = 16.54 ± 2.77-mmHg) and GAT measurements (IOP = 15.07 ± 2.35-mmHg), DCT measurements were significantly higher than NCT and GAT (p < 0.001). There was a significant correlation between CCT with both NCT (r = 0.260, p = 0.003) and GAT measurements (r = 0.257, p = 0.005). There was a weak correlation that was not statistically significant between CCT and DCT (r = 0.160, p = 0.079).Conclusion: The IOP measurements with DCT seem to be less dependent on CCT. NCT appears to be more affected by variation in CCT than GAT.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call