Abstract

We explored chromatographic conditions to obtain high resolution in protein separations by ion-exchange chromatography (IEC) on a macroporous anion-exchange resin of 10 μm in particle diameter. We studied effects of flow-rate, gradient time (steepness of salt concentration gradient) and column length on resolution in wide ranges. It was found that very high resolutions are attainable at long gradient times with long columns. The resolution continuously became higher as the gradient time and the column length became longer except in some special cases. The dependence of resolution on gradient time was particularly great when the column was long and the gradient time for the change of 0–0.5 M NaCl was longer than 2 h. On the other hand, the effect of flow-rate on resolution was very small. Although the separations at long gradient times with long columns have not been popular in high-performance IEC and it takes several hours for one separation, such separations should be advantageous when very high resolutions are required like in proteomics research.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.