Abstract

Even with extremely high output from a bipolar coagulator applied close to a vessel, the patency in the femoral arteries and veins of 20 rats was found to be excellent. Scanning electron microscopy and histopathologic studies showed minimal disturbances within the intima. When the extremely high pressure (350 gramforce/mm2) of the Ikuta microvascular occlusion clamp was applied to the rat femoral veins for 1 hour, only one out of six veins became nonpatent. Lower pressures had no detrimental effects on the other 33 veins. When high clamping pressure and high output bipolar coagulation were applied together in 20 femoral veins of the rat, patency rate was 80 percent. From the results, it appears that bipolar coagulation is extremely safe, while clamping can be harmful to patency in microvascular anastomoses.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.