Abstract
Currently, there is little understanding of the role of echogenic needles and beam steering at moderate angles of needle insertion. The ultrasound images of the echogenic and nonechogenic needles inserted into pork at 40°, 50°, and 60° were scored by anesthesiologists on a scale of 0-10. The effect of different levels of beam steer was also explored. At 40°, steep beam steering improves visualization of both nonechogenic and echogenic needles to an equal, satisfactory level. At 50° and 60°, visualization of nonechogenic needles is poor, whereas visibility of an echogenic needle was adequate and may be improved with steep beam steering.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.