Abstract

To study the effects of balloon angioplasty (PTA) of the femoropopliteal artery with intravascular ultrasound (IVUS). Corresponding IVUS cross-sections (n = 1033) obtained before and after PTA from 115 procedures were analysed. Vascular damage including plaque rupture, dissection and media rupture was assessed. Free lumen area (FLA), media-bounded area (MBA) and plaque area (PLA) were measured. After PTA vascular damage was seen at the target site in 83 (72%) arteries: plaque rupture in 30 (26%), dissection in 66 (57%) and media rupture in 20 (17%) arteries. The FLA increased from 5.4 +/- 3.4 mm2 to 14.1 +/- 5.0 mm2 (p < 0.001), MBA increased from 26.9 +/- 10.0 mm2 to 32.9 +/- 10.7 mm2 (p < 0.001) and PLA decreased from 21.6 +/- 8.5 mm2 to 18.8 +/- 8.0 mm2 (p < 0.001). The increase in MBA accounted for 68% of lumen gain. The frequency of vascular damage and the relative contribution of MBA increase and PLA decrease to luminal gain were not different in procedures with balloon diameter < or = 5 mm and > or = 6 mm. Vascular damage is common following PTA. Lumen gain is mainly due to vessel expansion and, to a lesser extent, to a decrease in plaque area.