Abstract

We quantified the level of backscatter radiation generated from physicians' heads using a phantom. We also evaluated the shielding rate of the protective eyewear and optimal placement of the eye-dedicated dosimeter (skin surface or behind the Pb-eyewear). We performed diagnostic X-rays of two head phantoms: Styrofoam (negligible backscatter radiation) and anthropomorphic (included backscatter radiation). Radiophotoluminescence glass dosimeters were used to measure the eye-lens dose, with or without 0.07-mm Pb-equivalent protective eyewear. We used tube voltages of 50, 65 and 80kV because the scattered radiation has a lower mean energy than the primary X-ray beam. The backscatter radiation accounted for 17.3-22.3% of the eye-lens dose, with the percentage increasing with increasing tube voltage. Furthermore, the shielding rate of the protective eyewear was overestimated, and the eye-lens dose was underestimated when the eye-dedicated dosimeter was placed behind the protective eyewear. We quantified the backscatter radiation generated from physicians' heads. To account for the effect of backscatter radiation, an anthropomorphic, rather than Styrofoam, phantom should be used. Close contact of the dosimeter with the skin surface is essential for accurate evaluation of backscatter radiation from physician's own heads. To assess the eye-lens dose accurately, the dosimeter should be placed near the eye. If the dosimeter is placed behind the lens of the protective eyewear, we recommend using a backscatter radiation calibration factor of 1.2-1.3.