Abstract

To evaluate different modes of cementless fixation of hemispherical cups, we operated on 87 hips in 81 patients using 4 different means of cup fixation. The hips were randomly assigned to fixation with press-fit technique only (PF), or with augmentation with screws (S), pegs (P), or hydroxyapatite (HA) coating. The patients were evaluated with radiostereometric analysis (RSA) for cup migration and wear, conventional radiography for osteolysis, and Harris Hip Score for clinical outcome over 5 years. The fixation of the cups did not differ between the groups, but HA showed a tendency to decrease proximal migration. HA-coated cups displayed the best interface with hardly any signs of radiolucent lines, indicating a superior sealing effect of the HA coating. Cups with screws or pegs had more radiolucent lines and osteolytic lesions than the other groups. Radiolucent lines were correlated to higher proximal migration, young age, and female gender (r2=.2). The wear rate of the ethylene oxide-sterilized polyethylene liner was high (0.2 mm/y) but did not differ between the groups. Two cups with a perioperative fracture of the acetabular rim showed large initial migration but stabilized thereafter.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.