Abstract

The effect of pyrocoated graphite, uncoated graphite, metal-carbide, and metal atomization surfaces on the quantitation of molybdenum and ytterbium by electrothermal atomization atomic absorption spectrometry was investigated. The peak shape was affected by heating rate and the different surfaces gave different shapes. Except for the case of uncoated graphite, the sensitivities and detection limits were similar for all surfaces. In a sodium chloride matrix it is preferable to use uncoated graphite for molybdenum because an ashing stage greater than the boiling point of sodium chloride can be used without loss of molybdenum. Tube lifetime depended on atomization temperature, atomization time and the matrix.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.