Abstract

This study examined the effect of artificial lung compliance (C) on pulmonary system (PS) impedance and right ventricular function during in-series attachment of the MC3 Biolung in adult sheep. Compliances, C, of 0-20 ml/mm Hg were tested at the Biolung inlet. Results indicate the PS 0 harmonic input impedance modulus was not affected by C. The PS first harmonic input impedance modulus (Z1) was 10.9 +/- 3.2 mm Hg/(l/min) at C = 0 ml/mm Hg and minimized to 2.41 +/- 0.79 mm Hg/(l/min) at C > or = 0.5 ml/mm Hg. Cardiac output was 58% +/- 10% of its pre-Biolung attachment, baseline value at C = 0 ml/mm Hg and was maximized to an average of 75% +/- 11% at C > or = 0.5 ml/mm Hg. The left ventricular lateral-to-anteroposterior axis length ratio, which decreases with leftward septal shift, increased with C from 0.52 +/- 0.12 at C = 0 ml/mm Hg to 0.76 +/- 0.06 at C = 5 ml/mm Hg (p < 0.05), but decreased slightly with C at C > 5 ml/mm Hg. Therefore, the ideal C for right ventricular function is at least 0.5 ml/mm Hg and may be as high as 5 ml/mm Hg to minimize septal shift.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.