Abstract

Apodization is a generally applied signal processing method in FT-ICR MS (Fourier transform ion cyclotron resonance mass spectrometry) analysis for complex petroleum samples that is used to lower the spectral leakage and smoothen the spectrum line shape. Present study evaluates the effect of apodization on the FT-ICR MS results through a systematic examination on both simulated and actual heavy petroleum signals. The result shows that the resolving power, instead of the dynamic range, is the key factor for heavy petroleum FT-ICR MS analysis due to the severe peak overlapping. Window functions with high dynamic range will reduce the resolving power so that many of the species cannot be detected. The improvement on accuracy of apodization with peak intensity is negligible and all the window functions have similar intensity error. Low dynamic range window functions perform better for FT-ICR MS analysis since they have less resolving power reduction and the spectral leakage is eliminated to ensure correct peak selection. Sine-bell and low beta value Kaiser window functions are recommended for complex mixtures.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.