Abstract

Although the automatic implantable cardioverter defibrillator (AICD) is effective against malignant ventricular arrhythmias, the effects of AICD patches on left ventricular diastolic properties have not been defined. Accordingly, extrapericardial (group E, n = 5) or intrapericardial (group I, n = 6) AICD patches were implanted through a median sternotomy in 11 anesthetized pigs. Six weeks later, using a left thoracotomy, the hearts were arrested with hypothermic cardioplegia. A balloon catheter was inserted into the left ventricle through the aortic root, and pressure-volume curves were measured before and after sequential removal of patches and pericardium. A dense intrapericardial fibrotic reaction in group I was not present in group E. Normalized left ventricular filling volumes in group E were significantly larger at pressures of 5.1 to 10, 15.1 to 20, and 20.1 to 28 mm Hg compared with group I ( p < 0.05). We conclude that intrapericardial AICD patches adversely affect left ventricular diastolic pressure-volume relations and recommend that AICD patches be placed in the extrapericardial location clinically whenever possible.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.