Abstract

Accelerometer step count accuracy may be affected by gait speed, device positioning, and analysis algorithm selection. We assessed ActiGraph GT3X+ step count accuracy related to device placement and analysis algorithm in older adults with walking aids (n = 13) and without walking aids (n = 22). Participants (81.5 ± 5.0 years of age) completed a timed 100-m walk wearing five GT3X+ monitors (hips, ankles, lumbar spine). Individuals with walking aids had slower gait speeds (0.8 ± 0.20 m/s versus 1.2 ± 0.20 m/s without walking aids, p < .001). Intraclass correlation coefficient values for observed versus monitored steps were highest when ankle placement and the low frequency extension (LFE) algorithm were used (left ankle ICC = .989, right ankle ICC = .998). Using the GT3X+ ankle placement and analyzing data with the LFE algorithm resulted in the most accurate step counts in older adults.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.