Abstract

Electrical Impedance Tomography (EIT) is a low-cost imaging method which reconstructs two-dimensional cross-sectional images, visualising the impedance change within the thorax. However, the reconstruction of an EIT image is an ill-posed inverse problem. In addition, blurring, anatomical alignment, and reconstruction artefacts can hinder the interpretation of EIT images. In this contribution, we introduce a patient-specific structural prior mask into the EIT reconstruction process, with the aim of improving image interpretability. Such a prior mask ensures that only conductivity changes within the lung regions are reconstructed. To evaluate the influence of the introduced structural prior mask, we conducted numerical simulations with two scopes in terms of their different ventilation statuses and varying atelectasis scales. Quantitative analysis, including the reconstruction error and figures of merit, was applied in the evaluation procedure. The results show that the morphological structures of the lungs introduced by the mask are preserved in the EIT reconstructions and the reconstruction artefacts are decreased, reducing the reconstruction error by 25.9% and 17.7%, respectively, in the two EIT algorithms included in this contribution. The use of the structural prior mask conclusively improves the interpretability of the EIT images, which could facilitate better diagnosis and decision-making in clinical settings.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.