Abstract

The aim of this study was to investigate the effect of a macroscopic groove on bone response and implant stability during the early stages of healing using a rabbit tibia model. Anodized titanium implants with (n=24) and without (n=24) macroscopic grooves were prepared. A total of 12 rabbits were used and each received four implants: six rabbits with implants with macroscopic grooves (test group) and six rabbits with implants without macroscopic groves (control group). Histomorphometry, resonance frequency, and removal torque value were evaluated 2 and 6 weeks post-implant-insertion. At 2 and 6 weeks, there was no significant difference between the two groups in the percentage of bone-to-implant contact (P>0.05). At 6 weeks, the test group had significantly higher implant stability quotient values than the control group (P<0.05). At 2 and 6 weeks, implants with grooves showed a significantly greater resistance to reverse torque than control implants (P<0.05). The groove on the oxidized titanium surface may increase both resistance to shear load and adhesion at the bone-implant interface. A geometric feature such as a macroscopic groove may facilitate osseointegration and increase implant stability in various clinical conditions. Further studies are required to confirm whether the improvement in implant stability will enhance treatment success in humans.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.