Abstract

Objective: To determine the effect of the intensity of two light curing units on the biocompatibility, flexural strength and elastic modulus of a composite resin.   Methodology: Two groups of Filtek Z250XT (3M ESPE) composite resin were created, each one photopolymerized using different intensities (<400 mW/cm2 for 40s and> 800 mW/cm2 for 20s). Cell viability was analyzed by MTT assay at 24 and 48 hours following the ISO 10993-5 standard. The flexural strength and elastic modulus were analyzed following the ISO 4049 standard. Results: In the group photopolymerized with an intensity <400 mW/cm2, cytotoxicity was statistically higher both at 24 and 48 hours and flexural strength and elastic modulus were statistically lower. Conclusion: A polymerization intensity <400 mW/cm2 increases the levels of cytotoxicity and decreases the mechanical properties of composite resins. The importance of the periodic control of the light curing units is emphasized.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call