Abstract

To explore differences in BNP concentration and stability and evaluate BNP accuracy in different collection tubes. BNP concentrations in heparin/glass, EDTA/glass, and EDTA/ PET tubes were measured on the Mindray CL-6000i at 0.5, 1, 2, and 4 h after collection. Differences were evaluated using Wilcoxon's paired tests and Bland-Altman plots. BNP stability and measurement accuracies were estimated using Kruskal-Wallis H tests and recovery tests. BNP concentrations in EDTA/glass tubes were 31.4% higher than those in heparin/glass tubes and 3.04% lower than those in EDTA/PET tubes. BNP stability significantly decreased in the heparin/glass tube. BNP remained stable in EDTA/glass and EDTA/PET tubes at room temperature for 4 h. BNP recovery rates in heparin/glass, EDTA/glass, and EDTA/PET tubes were 77.46, 86.04, and 88.23%, respectively. Plasma in EDTA/glass and EDTA/PET tubes is suitable for BNP measurement on the Mindray CL-6000i.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.