Abstract

The received electrical signal from a pulse-echo system insonifying a planar acoustical interface was measured for varying degrees of rms roughness (0-0.16 mm), angle of incidence (typically +/-7/spl deg/) and range to the transducer. A planar and a focused 5 MHz transducer was used. When insonifying a smooth interface, the normalized spectrum of the received signals for a planar transducer exhibits an increasing number of nulls with increased angle of insonification, as predicted from numerical modeling while the dependence on insonification angle for the focused transducer was smaller and the null pattern was much less distinct. For the planar transducer and for the focused transducer with the interface located at the geometrical point of focus, the energy of the received signal as a function of incident angle was approximately Gaussian with maximum at 0/spl deg/. For the smooth interface, the -3 dB width was /spl sim/0.5/spl deg/ and /spl sim/3/spl deg/ for the planar and the focused transducer, respectively.