Abstract
Purpose: The feasibility of hyperpolarized 129Xe for fast MR angiography (MRA) was evaluated using the echo-planar imaging (EPI) technique. Material and Methods: Hyperpolarized Xe gas was dissolved in ethanol, a carrier agent with high solubility for Xe (Ostwald solubility coefficient 2.5) and long relaxation times. The dissolved Xe was injected as a bolus into a flow phantom where the mean flow velocity was 15 cm/s. Ultrafast EPI images with 44 ms scan time were acquired of the flowing bolus and the signal-to-noise ratios (SNR) were measured. Results: The relaxation times of hyperpolarized Xe in ethanol were measured to T1=160±11 s and T2≈20 s. The resulting images of the flowing liquid were of reasonable quality and had an SNR of about 70. Conclusion: Based on the SNR of the obtained Xe EPI images, it was estimated that rapid in vivo MRA with 129Xe may be feasible, provided that an efficient, biologically acceptable carrier for Xe can be found and polarization levels of more than 25% can be achieved in isotopically enriched 129Xe.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.