Abstract
PurposeTo evaluate diffusion-weighted imaging (DWI) using echo planar imaging (EPI) with compressed SENSE (EPICS) of the head and neck magnetic resonance imaging (MRI). MethodWe retrospectively observed 32 patients who underwent head and neck DWI according to either the conventional method (SENSE, reduction factor = 2), fast scanning method (SENSE, reduction factor = 4), or fast scanning method with EPICS (EPICS, reduction factor = 4). For quantitative analysis, contrast-to-noise-ratio (CNR), apparent diffusion coefficient (ADC) values, geometric distortion, and coefficient of variations (CV) were measured and compared. For qualitative analysis, all images were independently and blindly evaluated by two board-certified radiologists. ResultsEPICS revealed the higher CNR between all location compared to those of SENSE with reduction factor = 4. Distortion in the anterior-posterior direction was significantly lower on EPICS than on the conventional scan (p = 0.02). A comparison between the ADC values of the EPICS and conventional scan revealed no significant differences. The CV was significantly lower for EPICS than the conventional scan [DWI: 0.22 (IQR: 0.15–0.30) vs 0.32 (IQR: 0.24–0.40), p = 0.02]. ConclusionsCompressed SENSE combined with the high acceleration factor can improve image quality, homogeneity, and distortion in the head and neck DWI maintaining ADC values and the scan time duration.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.