Abstract
The feasibility of BOLD fMRI in association with electrical somatosensory stimulation on spontaneously breathing, isoflurane-anesthetized mice was investigated using spin-echo, echo-planar imaging (EPI) on a vertical narrow-bore 9.4 T magnet. Three experiments were performed to derive an optimal fMRI protocol. In Experiment 1 (n = 9), spin-echo BOLD responses to 10% CO2 challenge under graded isoflurane (0.25-1.25%) ranged from 10 +/- 2% to 3.5 +/- 0.9%; the optimal BOLD contrast-to-noise ratio peaked at 0.75% isoflurane. In Experiment 2 (n = 6), hindpaw somatosensory stimulations using 1-7 mA under 0.75% isoflurane revealed the optimal BOLD response was at 6 mA. In Experiment 3 (n = 5), BOLD responses to 4 and 6 mA stimulation under 0.75% and 1% isoflurane were evaluated in detail, confirming the optimal conditions in Experiment 2. These results demonstrated that BOLD fMRI using single-shot, spin-echo EPI in a mouse somatosensory stimulation model could be routinely performed on high-field, vertical, narrow-bore magnets. This protocol might prove useful for fMRI studies of transgenic mice.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.