Abstract

Accurate detection of QRS complexes is essential for the investigation of heart rate variability. Several transform techniques have been proposed and extensively used for the detection and analysis of QRS complexes. In this proposed work, the de-noised ECG signal is subjected to a modified S-transform for QRS complex detection.The performance analysis of the proposed work is evaluated using parameters such as sensitivity, positive predictivity and accuracy. The algorithm delivers sensitivity, positive predictivity and overall accuracy of 99.91, 99.91 and 99.77%, respectively. Furthermore, a search back mechanism is employed, which specifies the filtered electrocardiogram (ECG) segment, which was traced for the true R-peak locations. The modified S-transform based QRS complex detection algorithm provides an excellent search back range of only ±2 samples in comparison with other earlier proposed algorithms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.