Abstract

Non-linear loading of the utilities supply introduced significant electromagnetic field (EMF) interference severe enough to disrupt electrocardiograph (ECG) monitoring and recordings in the new Emergency Department (ED) at the Royal Melbourne Hospital (RMH). This interference was evident even though standard Mu-metal shielding had been installed over the main hospital power feed which runs underneath the department. Investigations revealed that the source of the interference was due to 3rd harmonic currents flowing in the mains cable. This interference was suppressed by introducing third harmonic current into the main power cable in anti-phase to the interfering signal.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.