Abstract

In this study, a new compression algorithm for ECG signal is proposed based on selecting important subbands of wavelet packet transform (WPT) and applying subband-dependent quantization algorithm. To this end, first WPT was applied on ECG signal and then more important subbands are selected according to their Shannon entropy. In the next step, content-based quantization and denoising method are applied to the coefficients of the selected subbands. Finally, arithmetic coding is employed to produce compressed data. The performance of the proposed compression method is evaluated using compression rate (CR), percentage root-mean-square difference (PRD) as signal distortion, and wavelet energy-based diagnostic distortion (WEDD) as diagnostic distortion measures on MIT-BIH Arrhythmia database. The average CR of the proposed method is 29.1, its average PRD is <2.9 % and WEDD is <3.2 %. These results demonstrated that the proposed method has a good performance compared to the state-of-the-art compression algorithms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.