Abstract

The continuous real time reduction of baseline wander is a considerable problem in electrocardiography during exercises. Our solution consists of spectral filtering. The legitimacy of high-pass filtering of the ECG by means of digital linear phase filters with a low cut-off frequency as high as the heart rate is shown. The specifications of these filters are derived from experimental results. Special hardware is presented that simultaneously performs the desired real-time filter operation in four ECG leads.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.