Abstract

ObjectivesEfficient interruption of Ebola virus disease (EVD) transmission chains critically depends on reliable and fast laboratory diagnosis. We evaluated the performance of the EBOLA Virus Antigen Detection K-SeT (EBOLA Ag K-SeT), a new rapid diagnostic antigen test in field settings. MethodsThe study was conducted in a field laboratory located in Freetown (Sierra Leone) by the Italian National Institute for Infectious Diseases ‘L. Spallanzani’ and the EMERGENCY Onlus NGO. The EBOLA Ag K-SeT was tested on 210 residual plasma samples (EVD prevalence 50%) from patients hospitalized at the EMERGENCY Ebola treatment center in Goderich (Freetown), comparing the results with quantitative real-time PCR. ResultsOverall, the sensitivity of EBOLA Ag K-SeT was 88.6% (95% confidence interval (CI), 82.5–94.7), and the corresponding specificity was 98.1% (95% CI, 95.5–100.7). The positive and negative predictive values were 97.9% (95% CI, 95.0–100.8) and 89.6% (95% CI, 84–95.2), respectively. The sensitivity strongly increased up to 98.7% (95% CI, 96.1–101.2) for those samples with high virus load (≥6.2 log RNA copies/mL). ConclusionsOur results suggest that EBOLA Ag K-SeT could represent a new effective diagnostic tool for EVD, meeting a need for resource-poor settings and rapid diagnosis for individuals with suspected EVD.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.