Abstract

Skin graft has been known to be prone to failure. This study was aimed to make a simplification of the negative pressure wound therapy (NPWT), which follows EASEPort (effective, affordable, safe, easily handled, and portable) concept to support the take of skin graft. The design of the EASEPort-NPWT was then made and technically verified. Thereafter, an animal experimental study comparing the EASEPort-NPWT to the classic tie-over technique on skin graft over exudative wound was conducted. The EASEPort-NPWT was verified to be able to yield and sustain the subatmospheric pressure needed. In the animal study, the treatment group showed better skin graft survival rate (97.55 ± 11.18% take) than the control group (54.88 ± 19.73%) on day-7. Histopathology examination showed good quality of the skin structures taken from the treatment group, which was better than the structures of the skin in the control group. In summary, this study has been able to fulfill its objective to create a device following EASEPort concept. Subsequently, the EASEPort-NPWT was able to enhance skin graft survival rate in exudative wound.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.