Abstract
This study evaluated two automated external defibrillators (AEDs) to determine the extent to which different AED designs affect the performance of older adults over the age of 65 years. Forty-two untrained laypersons (20 young adults, M = 27 years old and 22 older adults, M = 71 years old) were randomly assigned to use training versions of either the Medtronic LifePak CR Plus or the Zoll AED Plus during a simulated cardiac arrest using an iStan patient simulator. A number of aspects of AED designs were difficult for the two age groups. Fifty percent of older users and 55% of younger users were able to use the AEDs to deliver a shock to the patient. For those who were able to complete the appropriate tasks steps, younger adults ( M = 109.90, SD = 23.30) delivered the initial shock faster than older adults ( M = 141.46, SD = 50.62). Task completion rates differed by age group and AED type. The most problematic steps across age groups included opening and activating the AED, removing the liner and placing the pads on the patient simulator. This paper will also discuss the implications of the AED evaluations and redesign recommendations.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Proceedings of the Human Factors and Ergonomics Society Annual Meeting
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.