Abstract

Early onset prosthetic valve stenosis is an uncommon complication after valve replacement surgery and is often caused by thrombus formation. Frequently it can be diagnosed by echocardiography and managed with optimizing anticoagulation and/or thrombolysis. We review a unique case of early bi-leaflet mechanical heart valve (BMHV) dysfunction where the patient rapidly progressed to cardiogenic shock requiring emergent re-do mitral valve surgery. Intraoperatively, the valve leaflets were found to be almost completely immobile secondary to thrombus formation directly on the hinges of the valve. This case demonstrates how the leaflet orientation of a BMHV affects transmitral flow and fluid dynamics. Furthermore, we also discuss left atrial vortex formation in the setting of atrial fibrillation, kinetic energy transfer through an anatomically implanted mechanical mitral valve, and their roles in contributing to early prosthetic valve thrombosis despite adequate anticoagulation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.