Abstract

Endovascular aortic arch repair using multibranch devices has been applied in patients considered at high risk for open repair. The aim of this case series was to report the early outcomes in patients managed with a new design 3 branch arch custom-made device, including a retrograde left common carotid artery (LCCA) branch. The Preferred Reporting Of CaSe Series in Surgery (PROCESS) guidelines were followed. All consecutive patients undergoing endovascular repair of an aortic arch lesion with a custom-made triple-branch device, including a retrograde LCCA branch (Cook Medical, Bloomington, IN, USA), between October 27, 2022, and February 28, 2023, were included. The presence of an arch aneurysm (degenerative or post-dissection) with diameter ≥55 mm and high risk for a conventional open repair set the indication for treatment. The primary outcomes were technical success and mortality at 30 days. Early morbidity and reinterventions were considered as secondary outcomes. Eight elective patients (87.5% men, mean age 72.3±27.0 years) were included. Five of them (62.5%) had undergone a previous ascending aorta repair of an acute type A aortic dissection. All patients were asymptomatic, except one, with left recurrent laryngeal nerve compression. The mean maximum aortic diameter was 70.4±21.0 mm. Percutaneous femoral and axillary access was used in all cases except three in which a cut down for right carotid access was performed. Technical success was 100%. Femoral access to the LCCA and implantation of the bridging stent was performed without technical challenges. No death nor cerebrovascular event was recorded during the 30 day follow-up. Five patients (62.5%) presented major complications, 3 related to access needing reintervention and the remaining related to congestive heart failure (CHF), which were managed successfully with medical treatment. Follow-up (range 1-4 month) was uneventful, except for one patient who presented a secondary type Ia endoleak. According to our early experience, the presence of a retrograde branch facilitated the revascularization of the LCCA through femoral access, decreasing the risk of cerebrovascular morbidity. Further analyses with longer follow-up are needed to evaluate the safety and efficacy of the device. Data arising mainly from the retrograde branch for the revascularization of the LSA are encouraging from a variety of devices. The premiminary experience with a triple-branched arch device, with a retrograde branche for the LSA but also for the LCCA, was associated with no 30 day mortality and 100% technical success.The device's design allowed swift catheterization and completion of the LCCA revascularization using femoral access exclusively.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call